beautypg.com

Signal analysis, Alarm generation – EXFO IQS-8100 Series Transport Blazer for IQS-600 User Manual

Page 153

background image

Port Tabs

SONET/SDH Application

139

Port TX (Electrical Interfaces)

³

LBO (Line Build Out): The LBO allows to meet the interface
requirements over the full range of cable lengths. LBO is not available
with E1, E2, E3, and E4 interfaces.

For DS1:
Preamplification values: +3.0 dBdsx (533-655 ft), +2.4 dBdsx
(399-533 ft)
, +1.8 dBdsx (266-399 ft), +1.2 dBdsx (133-266 ft), and
+0.6 dBdsx (0-133 ft).
Cable simulation (CSU Emulation mode) values: 0.0 dBdsx, -7.5
dBdsx
, -15.0 dBdsx, and -22.5 dBdsx.

For DS3: 0 to 225 feet range, 225 to 450 feet range, and Cable
Simulation 900 ft)
.

For STS-1e/STM-0e: 0 to 225 feet range, 225 to 450 feet range, and
Cable Simulation 900 ft).

For STS-3e/STM-1e: 0 to 225 ft.

Signal Analysis

Output Presence: Indicates the presence of a signal at the output port
(green) or not (gray).

Alarm Generation

³

Type

LOS (Loss Of Signal): Turns off the output port signal.

³

On/Off button: Press On/Off to enable/disable the alarm generation.