J1 trace – EXFO IQS-8100 Series Transport Blazer for IQS-600 User Manual
Page 357
SDH Tabs
SONET/SDH Application
343
HOP RX (SDH)
³
ERDI-CD (Enhanced RDI - Connectivity Defect): The ERDI-CD alarm is
declared when bits 5, 6 and 7 of the G1 byte contain the “110” pattern
in five consecutive frames.
³
ERDI-PD (Enhanced RDI - Payload Defect): The ERDI-PD alarm is
declared when bits 5, 6 and 7 of the G1 byte contain the “010” pattern
in five consecutive frames.
J1 Trace
³
Received Message: Displays the J1 value in 16-bytes or 64-bytes R > and F >, represent respectively a carriage return and a line feed. ³ Enable HP-TIM: Enables the Trace Identifier Mismatch for the ³ Expected Message: Allows entering the message that is expected. J1 ³ Expected Format: Allows the selection of the format expected.
format. The
last two bytes of a 64-bytes format,
expected message defined when the Enable HP-TIM check box is
selected. The Enable HP-TIM check box has to be selected to give
access to the expected trace format and message.When the Enable
HP-TIM check box is cleared, the J1 1-byte is available from the HOP
OH RX (SDH) on page 344.
value should be ASCII suitable characters. The default message is
EXFO SONET/SDH for 16 bytes and EXFO SONET/SDH Analyzer high
order path trace test message for 64 bytes. However, with VCAT/LCAS
the default message will be EXFO followed by the VCG number (VCAT
and LCAS) and the SQ (VCAT only) number (for example
EXFO-VCG1-SQ0) for both 16 and 64 bytes formats.
Choices are 16 or 64 bytes. The default setting is 16 bytes.