Creating and Starting a Test Case

86

IQS-8100 Series Transport Blazer

Typical Test Cases

4f. For GFP, select the UPI (User Payload Identifier),

EXI (Extension Header Identifier), and CID
(Channel IDentifier) parameters. GFP is
available in Coupled test mode only. Refer to
GFP Frame TX on page 432 for UPI and EXI, and
to GFP Channel TX on page 435 for CID.

4g. For External Ethernet, which is available with

GFP, select the interface and its rate. Refer to
GFP Client TX on page 444 for more information.
Go to step 8.

4h. For Pattern, go to step 5.

4i. Press Next or Finish.

4j. Repeat step 4 as required to complete the path/mapping. See

Supported Paths/Mappings on page 59 for more information.

5. Pattern Configuration:

5a. Set the pattern parameters. Refer to

Pattern TX on page 405 and Pattern
RX
on page 409 for more
information.

6. Press Finish to complete the test setup.

The Grid tab closes and automatically
switch to the Alarm summary tab.

7. For additional configuration parameters

and results, refer to Summary Tabs on page 125, Port Tabs on page 137,
SONET Tabs on page 211, DSn Tabs on page 269, SDH Tabs on
page 303,
PDH Tabs on page 369, BERT Tabs on page 405, Advanced
Tabs
on page 411
, and Common Tabs on page 491. For Unframed
mode, only Summary, Port, and BERT tabs are available.

8. Press the Start button to start the test. Refer to Global Test Status and

Controls on page 35 for more information on test management.