OTN Tabs

208

IQS-8100 Series Transport Blazer

GMP TX

GMP TX

Generic Mapping Procedure (GMP) is available with GFP-F over ODUflex,
GFP-T over ODU0, and SONET/SDH over ODU0. GMP is available on both
HO and LO for ODU0 over ODU2 or ODU0 over ODU3 mapping.

Press TEST, ODUk, and GMP (under ODUk TX).

Configuration

Note: Only available for Ethernet in ODUflex over ODU3. Not supported with

ODUflex CBR test case. This value is set to 15230 for Ethernet in ODUflex
over ODU2.

Cm corresponds to the number of payload bytes per framed transported.
Choices are from 15165 to 15230 bytes. The default setting is 15230 bytes.
A minimum Cm value higher than 15165 may be imposed in order to
guarantee that no packet is lost.

Statistics

Cm Min and Cm Max: Indicates respectively the minimum and maximum
Cm values transmitted during the test.

CnD Min and CnD Max: Indicates respectively the minimum and
maximum CnD values transmitted during the test.