Phy alarm generation, Mac error injection – EXFO IQS-8100 Series Transport Blazer for IQS-600 User Manual
Page 411
![background image](/manuals/736526/411/background.png)
Ethernet Tabs
SONET/SDH Application
397
Error/Alarm TX
PHY Alarm Generation
³
Type: The following alarms are available:
³
On/Off button: The On/Off button is used to activate/deactivate the
selected alarm. This setting is disabled (Off) by default.
MAC Error Injection
³
Type: The following error is available with both manual and automated
injection modes: FCS.
³
Amount: Allows the selection of the amount of manual error to be
generated. Choices are 1 through 50. The default setting is 1.
³
Send button: Press Send to manually generate error(s) according to
the error type and the amount of selected error.
³
Rate: Press the Rate field to select the rate for the automated error.
Choices are: 1.0E-02, 1.0E-03, 1.0E-04, 1.0E-05, 1.0E-06, 1.0E-07,
1.0E-08, 1.0E-09 or user definable from 1.0E-09 to 1.0E-02. The default
setting is 1.0E-04.
Type
Avaialble with
Ethernet Gb Ethernet 10G Ethernet
Local Fault Generates a local
fault sequence.
X
X
Remote Fault: Generates a
remote fault sequence.
X
X
X
Link Down: Generates a
continuous PCS error (block
error). Not available with 10G
Ethernet mapped in GFP.
X