Odu tx, Error injection – EXFO IQS-8100 Series Transport Blazer for IQS-600 User Manual
Page 194
OTN Tabs
180
IQS-8100 Series Transport Blazer
ODU TX
ODU TX
Press TEST, ODUk, and ODU (under ODUk TX).
Error Injection
Allows manual or automated error injection.
³
Type: The following error types are available with both manual and
automated injection mode: ODU-BIP-8, and ODU-BEI. The default
setting is ODU-BIP-8.
³
Amount: Select the amount of errors to be generated. Choices are 1
through 50. The default setting is 1.
³
Send button: Press Send to manually generate error(s) according to
the Error Type and the Amount of Errors selected.
³
Rate: Press Rate to select the injection rate for the selected error. The
rate must be within the minimum and maximum values specified. The
default setting is 6.5E-05.
³
Continuous: Generates the selected error to its theoretical maximum
rate when the Continuous check box is selected. The Continuous
check box is cleared by default.
³
On/Off button: The On/Off button is used to activate/deactivate the
selected automated error at the rate specified or at its theoretical
maximum rate when the Continuous check box is selected. This
setting is disabled (Off) by default.