Alarm analysis, Alarm generation – EXFO IQS-8100 Series Transport Blazer for IQS-600 User Manual

Page 499

background image

Next-Generation Tabs

SONET/SDH Application

485

LCAS Sink

Alarm Analysis

³

TLCR (Total Loss of Capacity Receive): The TLCR alarm is declared
when the number of active member equal zero while the number of
provisioned member is bigger than zero.

³

PLCR (Partial Loss of Capacity Receive): The PLCR alarm is declared
when the PLCR Threshold is reached while both numbers of active
member and provisioned member are bigger than zero.

³

FOPR (Failure of Protocol Receive): The FOPR alarm is declared when
persistent CRC or SQNC is present.

³

SQNC (Sequence Indicator Non-Consistent): The SQNC alarm is
declared when the members that carry the NORM, DNU or EOS
message do not have a unique sequence indicator.

Note: Refer to Alarm/Error Measurements on page 47 for H/C LEDs and Seconds

information.

Alarm Generation

³

Type

UMST (Unexpected Member Status): The UMST is generated by
constantly forcing a MST = OK for the selected SQ that is not currently
used in the VCG.

³

SQ (Sequence Indicator): Allows selecting the SQ number for the
selected alarm to be generated. Possible values are from 0 to 63 for
LOP or 0 to 255 for HOP. The default setting is 1.

Note: Only SQ members not currently used in the VCG are available.

³

On/Off button: The On/Off button is used to activate/deactivate the
selected alarm. This setting is disabled (Off) by default.