4.3 Exploratory curves
797 VA Computrace – Software
79
Options
General parameters for peak evaluation.
Manual
Manual peak evaluation. The base points for base-
line evaluation must be set manually.
Automatic
Automatic peak evaluation. The base points for
baseline evaluation are evaluated automatically.
Reverse peak
Enable peak evaluation of reverse peaks (peaks
with opposite direction compared to the sweep
direction: negative peaks with anodic sweeps;
positive peaks with cathodic sweeps).
Reverse sweep
Enable peak evaluation of the reverse sweep
of
cyclic voltammograms (only available with CV and
CVS)
.
Minimum peak width (V.step
)
[ ≥ 0 ; 10 ]
Minimum peak width for peak recognition by
number of
Voltage steps
(= number of measure-
ment points).
Minimum peak height (A) [ > 50 pA ; 100 pA ]
Minimum peak height for peak recognition.
Smooth factor [ 1...6 ; 3 ]
Smoothing power for the Savitzky/Golay smooth-
ing of the baseline (
1
= minimum smoothing,
6
=
maximum smoothing).