3 General settings for exploratory and determination mode

797 VA Computrace – Software

52

U

t

Deposition potential

Deposition time

Potential limit

Maximum time

dt/dU

U

Peak voltage

Peak
area

Sweep parameters:

Hydrodynamic (measurement) [ on, off ; off ]

Enable/disable stirring of the RDE/SSE during the sweep.

Potential limit (V)

[ -5...+5 V ; -0.2 V ]

Voltage limit for PSA sweep.

Maximum time (s)

[ > 0 ; 20 s ]

Time limit for PSA sweep.

Stripping current (A) [ > 4e-009 ; 1e-005 ]

Constant current, applied for stripping after switching off the
potential.

Notes:
• The CCPSA measurement mode should only be used with RDE

electrodes (mainly with Hg film).