7.2 Dosing Device control
797 VA Computrace – Software
199
7.2
Dosing Device control
Dosing Device control selection
MAIN WINDOW
/ Utility / Dosino control
Start manual control of the Dosing Devices (possi-
ble: 700/800 Dosino , 685/805 Dosimat) con-
nected to the 797 VA Computrace stand
or the
846 Dosing Interface
.
Dosino control window
The
DOSINO CONTROL
window serves for manual control of the
Dosing Devices connected to the 797 VA Computrace stand
(under
Dosing Processor
) or the 846 Dosing Interface (under
Dosing Inter-
face
).
Dosino 1..7 (x mL)
3 Dosing Devices can be connected to the 797 VA
Computrace stand, 4 more to a 846 Dosing In-
terface. Each can be controlled individually.
Status [ read only ]
Display of the current Dosing Device status.
Total dosed volume (mL) [ > 0.01 mL ; 0 mL ]
Display of current volume dispensed since last fil-
ling (first value) and accumulated volume if the
display is not reset after filling (value in brackets).