Glp wizard – Metrohm 797 VA Computrace User Manual
Page 42
![background image](/manuals/587121/42/background.png)
2 Main window
797 VA Computrace – Software
30
Define, how long before the expiring date a war-
ning message is displayed.
Action, if validation expires
You can choose between just showing a warning
Show warning only
, or stopping the measure-
ments
Stop measurements
.
Note: GLP can only be started if there is no “blank user”. “Blank
user” has to be removed to apply GLP.
GLP Wizard
Clicking the
button on the
GLP
tab of the
GENERAL
SETTINGS
window starts the GLP Wizard. The GLP Wizard window
opens, and you can check which parts you want to execute.
Diagnostics
Performs several tests of the hardware of your in-
strument. You can choose which test should be
performed in the
DIAGNOSTICS
window.
Note: Occasional purging and stirring tests should
be done as “Check the purging”(see section
8.10 Check the purging) or “Check the stirring”
(see section 8.10 Check the stirring), and not with
the GLP Wizard (not to overwrite GLP data). It is
also possible to do the whole Diagnostics proce-
dure outside of the GLP Wizard: Click
DIAGNOS-
TICS
in
MAIN WINDOW /
HELP
to open the
DIAG-
NOSTICS
window.
Dummy cell tests
Performs an electronic validation of the 797 VA
Computrace.
Note: Occasional linearity and peak tests should
be done as “Perform a linearity test with the
dummy cell”(see section 8.10 Perform a linearity
test with the dummy cell) or “Perform a peak
test with the dummy cell” (see section 8.10
Perform a peak test with the dummy cell), and
not with the GLP Wizard (not to overwrite GLP
data).
Electrode test
Performs a validation of all three electrodes of the
797 VA Computrace.