Determination – Metrohm 797 VA Computrace User Manual
Page 102

5 Determination mode
797 VA Computrace – Software
90
Addition [ Manual, Automatic ; Manual ]
Selection of manual or automatic standard addi-
tion or recording of calibration curves:
Manual
Manual standard addition resp. recording of cali-
bration curves using a pipette.
Automatic
Automatic standard addition resp. recording of
calibration curves using one or several Dosing De-
vices.
Electrode
Selection of electrode (see Electrodes, section
3.1).
Drop size
Drop size for SMDE or HMDE (see Electrodes, sec-
tion 3.1).
Stirrer
Stirrer settings (see Stirring, section 3.4).
Dosing Device settings
(see Dosing Devices
, sec-
tion 5.2
).
Potentiostat settings (see Potentiostat, section
3.3).
Edit working method parameters (see
Determination, Voltammetric, Substances,
Calculations and Documentation).
Determination
The
Determination
tab of the
EDIT WORKING METHOD PARAME-
TERS
window
contains general specifications for performing the
determination. The parameters displayed depend on the selected
Calibration
technique and measurement
Technique
.