5.2 Working method specifications
797 VA Computrace – Software
107
Append
If this option is activated, the new XML -file is ap-
pended to a (possibly) already existing CSV-file.
Overwrite
If this option is activated, the new XML -file over-
writes a (possibly) already existing CSV-file.
Export determination to Auto Database
If this option is activated, the determintion data is
exported to a database.
Database
Database in which the data is to be exported.
The
database can be selected by clicking the button
.
Dosing Devices
The automatic use of Dosing Devices (possible: 700/800 Dosino,
685/805 Dosimat) for the addition of solutions has to be defined in
the
DOSINOS
window that is opened by clicking the
button in
WORKING METHOD SPECIFICATIONS
window.
Note: Make sure that
Automatic
is selected for
Addition
on the
WORKING METHOD SPECIFICATIONS
window.
Achtung: Dosino 1..3 refer to Dosinos connected to the 797 VA
Computrace. Dosino 4...7 refer to Dosinos connected to the 846
Dosing Interface (to MSB 1...4).
Use [ on, off ; off ]
Checkbox for
Dosing Devices
used for automatic
addition of solutions. For the addition of standard
solutions, the
Dosing Devices
must have the same
number as the
Standard solution No.
entered on
the
Substances
tab of the
EDIT WORKING
METHOD PARAMETERS
window.