Metrohm 797 VA Computrace User Manual
Page 192

6 Electroplating Bath VA
797 VA Computrace – Software
180
Add production bath to electrolyte [ on, off ; off ]
If is activated, the Production bath solution is ad-
ded to the Electrolyte solution. If it is 
not
activa-
ted, the Electrolyte solution is siphoned off before 
the Production bath solution is placed into the the 
measuring vessel. 
Initial electrode conditioning
(
see Initial electrode conditioning,
section 6.3
)
Check the checkbox to use conditioning.
No. of conditioning measurements [ 1...100 ; 1 ]
Number of replications of conditioning measure-
ments. The “total number of conditioning measu-
rements” is the 
No. of conditioning measurements
multiplied with the number of
Save last .. sweeps
(
Voltammetric
tab). “Total number of conditioning
measurements” must not exceed 100.
or
Auto Std.dev (%)[ >0.1 ; 1 ]
Define until which „Std.dev.“ the conditioning is 
continued. The „Std.dev.“ is calculated from all sa-
ved sweeps of the last two “conditioning measu-
rements”. 
Note: If you set a small value (e.g. < 1%), it can 
take a long time until reaching the target. 
Note: If both
No. of conditioning measurements
[
1...100 ; 1 ] and
Auto Std.dev
(%) [ >0.1 ; 1 ] are
activated, the conditioning is finished as soon as 
the first of the two criterias is reached. 
Additional conditioning after sample transfer
If this option is activated, an additional electrode 
conditioning is started after sample addition. This 
option can only be activated, if 
Initial electrode
conditioning
is activated.
Response curve [ path + file name ; ]
Choose the file which contains the desired respon-
se curve (the response curve can be recorded with 
the 
Calibration
technique "RC Record response
curve").
Achtung: To make sure that for "Suppressor ana-
lysis with 838 Advanced Sample Processor and 
RC" always the latest response curve file is taken 
for the calculation, the name of the file chosen for 
Response curve
, must match with the parameter
Sample identifier
on the Determination tab with
Calibration
technique "RC Record response curve".
