beautypg.com

Curves with high noise, Sqw problems – Metrohm 797 VA Computrace User Manual

Page 275

background image

9.6 Voltammetric problems

797 VA Computrace – Software

263

3. Exchange the RDE.
4. Has the electrode been conditioned (e.g. by using

Conditioning

cycles

and

Cleaning potential

) ?

5. If the concentration to be determined is considerably higher

than expected: reduce sample volume.

6. The background current is normally higher if RDE is used in

place of MME; a background current of several 100 nA is pos-
sible.

Curves with high noise

Did you work in the SqW mode (if yes, see SqW Problems, section
9.6
)? If not:
With all MME types:
1. Adjust sealing needle.
2. Change capillary and replace sealing needle (see Hardware

Manual). If necessary, clean the MME (see Hardware Manual).

3. Check tapping mechanism on stand. If tapping strength is too

weak, turn corresponding slotted screw on the valve block dur-
ing operation in an anticlockwise direction until a drop falls
each time the tapper is triggered (see Hardware Manual).

4. Electrolyte solution too old: make up a new one. Its working

life with organic additives may be as short as 1 day or less.

With DME/SMDE:
1. If the electrode drops irregularly: check the MME.

Adjust sealing needle. If necessary, change capillary

and

replace

sealing needle (see Hardware Manual).

2. If the electrode drops much too quickly: increase the

Voltage

step time

on the

Voltammetric

tab of the

EDIT WORKING MET-

HOD PARAMETERS

window.

With HMDE:
1.

If the electrode surface is

overloaded

: check deposition poten-

tial and time.

2. If no drops are at the capillary: change the capillary and replace

sealing needle (see Hardware Manual).

SqW Problems

In order to get good results for measurements in the “Square
wave” mode, the auto current range feature of the potentiostat
should not be used. The potentiostat parameters

Highest current

range

and

Lowest current range

should be set on the same level.

Finding the optimal current level needs trying. It should be as small
as possible, but larger than the expected maximum peak (to mini-
mize noise and not disturb the peak shape).