8.6 Analyze Electroplating Bath Solutions

797 VA Computrace – Software

227

General Settings in the 797 Software
Settings on the

Dosinos

tab of the

GENERAL SETTINGS

window in

the 797 Software:

It is recommended to do for all used Dosinos one "Prep cycle".

On the

Automation

tab, select the 838 Advanced Sample Processor.

The default settings can be adopted (click button <

Default

>).

Method Parameters for the 797
Set CVS or CPVS for

Mode

, and "MLAT" for

Calibration

.

Settings for the

DOSINOS

window:

Checking

Use for predose

for Dosino 1, VMS and additional Sup-

pressor concentrate are added to the measuring vessel before the
sample transfer. Together, they form the "Intercept solution". To
keep the Suppressor concentration stable, additional Suppressor
concentrate is added after the sample transfer.