8 How to ...?

797 VA Computrace – Software

244

Note: To make sure that always the latest "response curve file"
is taken for the calculation, the name of the "response curve fi-
le" defined for the parameter

Response curve

on the

Determina-

tion

tab (of the window

EDIT WORKING METHOD PARAMETERS

)

with

calibration

"RC Sample with response curve", must match

with the name defined for the parameter

Sample identifier

on

the

Determination

tab (of the window

EDIT WORKING METHOD

PARAMETERS

) with

Calibration

technique "RC Record response

curve".

Note: To make sure that always the latest "response curve file"
is taken for the calculation, the path of the calibration file defi-
ned for the parameter

Response curve

on the

Determination

tab

(of the window

EDIT WORKING METHOD PARAMETERS

with

calibration

"RC Sample with response curve", must match with

the path defined for the parameter

Data folder

(for the currently

logged-in user) on the tab

User Directories

of the window

USER

RIGHTS

.

Sample table
In the Sample table, list (according to the arrangement on the sam-
ple rack) for every empty vessel on the sample rack a method with

Calibration

technique RC Record response curve and for every sam-

ple a method with

Calibration

technique RC Sample with response

curve: