5.3 Monitor

797 VA Computrace – Software

111

overload. In this case, stop the measurement and change the
measurement parameters.
For a running voltage sweep there is a live display in the

MONITOR-

ING

window with automatic scaling of the axes. Manual rescaling

can be done by pressing the button or selecting the

Auto

scale

option of the context sensitive menu. At the end of each volt-

age sweep, the recorded curve is copied into the

DETERMINATION

CURVES

window.

Message windows during determination

For some steps in the operation sequence additional windows de-
manding an action or entry of the user appear.

PLACE BLANK

If the

Measure blank

option in the

Determination

tab is enabled, the message

Place blank and press

OK

appears. Add the blank solution into the meas-

uring vessel and press

.

PLACE SAMPLE

This window appears for all sample determina-
tions with standard addition or calibration curve.

The window displays date and time of the deter-
mination start and the title of the method used.
Date and time in the format

YYMMDDHHMM

(month-day-hour-minute) and the

Sample ID

are

used as default for the name of the determination
file to be saved automatically (e.g.

0706181712_sample.dth

).

Sample ID [ 32 characters ; "sample" ]

Identification for sample.

The expression defined

in the

Determination

tab is displayed and can be

changed if desired.

Sample amount [ > 0 ; 10 ]

Amount of sample added to the measuring vessel.
The value defined on the

Determination

tab is dis-

played and can be changed if desired.

Sample unit [ mL, g ; mL ]

Selection of unit for sample amount. The value de-
fined on the

Determination

tab is displayed and