beautypg.com

Perform a glp validation – Metrohm 797 VA Computrace User Manual

Page 268

background image

8 How to ...?

797 VA Computrace – Software

256

8. Enter

the

Sample ID

(used as part of the determination file

name) in the

PLACE SAMPLE

window and click

.

9. At the end of the measurement, a curve is printed out. This

curve should satisfy the following conditions:

• A symmetrical, gaussian-shaped peak should be plot-

ted. The evaluation must provide a result for the peak
voltage and the peak current, which are printed out
in the full report.

• The peak voltage E should be -450 … -550 mV.

Perform a GLP Validation

The GLP validation is performed with the

GLP Wizard

(see section

2.7, GLP Wizard) . To start it, click

on the GLP tab of

the

GENERAL SETTINGS

window.

.