3.2 VA measurement modes

797 VA Computrace – Software

47

Sweep parameters:

Hydrodynamic (measurement) [ on, off ; off ]

Enable/disable stirring of the RDE/SSE during the sweep.

Start potential (V)

[ -5...+5 V ; -0.9 V ]

Start voltage for the voltage sweep.

End potential (V)

[ -5...+5 V ; -0.1 V ]

Final voltage for the voltage sweep.

Base potential (V)

[ -5...+5 V ; -0.9 V ]

Base voltage for voltage sweep.

Pulse time (s)

[ ≥ 500 μs ; 0.04 s ]

Time interval during which an increasing voltage pulse is su-
perimposed on the base voltage.

Voltage step (V)

[ > 0 V ; 0.006 V ]

Voltage step for direct voltage ramp.

Voltage step time (s)

[ > 0 s ; 0.4 s ]

Time interval after which the voltage in the sweep is in-
creased or decreased by the amount

Voltage step

.

Sweep rate (V/s)

[ read only ]

Ramp slope calculated as

Voltage step / Voltage step time

.


Notes:

The NP measurement mode can be used with all types of elec-
trodes (but only in the exploratory mode!).

• The following conditions apply to the definition of the

Voltage

step time

:

Voltage step time > Pulse time + 10 ms

(HMDE/RDE)

Voltage step time > Pulse time + 220 ms

(DME)

Voltage step time > Pulse time + Drop size × 40 ms + 200 ms

(SMDE)

• The measurement time

t (i)

is defined as follows:

Pulse time 40 ms

t (i) = 20/16.67 ms

(

power frequency

50/60 Hz

)

Pulse time < 40 ms

t (i) = 0.5 × Pulse time