8 how to, 1 installation and program start, Install dosing devices for automatic addition – Metrohm 797 VA Computrace User Manual

Page 217: Switch on the instruments and start program, How to

background image

8.1 Installation and program start

797 VA Computrace – Software

205

8 How to ...?

8.1

Installation and program start

Install Dosing Devices for automatic addition

1. Connect Dosing Devices (possible: 700/800 Dosino , 685/805

Dosimat) to the 797 VA Computrace stand

or the 846 Dosing

Interface

(see Installation of Dosing Devices, section 1.3).

2. Make hardware settings for Dosing Devices.
3. Select

Automatic

for

Addition

in the

WORKING METHOD SPECI-

FICATIONS

window.

4. Define the addition or predose solution in the

DOSINOS

win-

dow (see Dosing Devices, section 5.2).

5. The number defined in the

No.

field of the

Substances

tab of

the

EDIT WORKING METHOD PARAMETERS

window should

correspond with the number of the Dosing Device used for the
addition of this substance.

Switch on the instruments and start program

1. Switch on PC.
2. Connect Dosing Devices ( see Installation of Dosing Devices,

section 1.3), Sample Changer (see Installation of 863 Compact
VA Autosampler
section 1.3, or Installation of 838 Advanced
Sample Processor
section 1.3) and the 843 Pump Station to the
797 VA Computrace stand.

3. Connect the 797 VA Computrace stand to the PC (via USB).
4. Switch on 797 VA Computrace stand, Sample Changer and 731

Relay Box.

5. Start 797 VA Computrace Software 1.3.x (see Starting the VA

Computrace program, section 2.2).

6. Enter

Name

and

Password

in the

VA COMPUTRACE LOGIN

win-

dow (see Login, section 2.6).

7. Select exploratory or determination mode (see Mode menu,

section 2.4).

8. Open the desired Exploratory mode windows (Exploratory

specifications, Exploratory curves) or Determination mode win-
dows (Working method specifications, Monitor, Determination
curves, Results, Sample table).