8 how to, 1 installation and program start, Install dosing devices for automatic addition – Metrohm 797 VA Computrace User Manual
Page 217: Switch on the instruments and start program, How to

8.1 Installation and program start
797 VA Computrace – Software
205
8 How to ...?
8.1
Installation and program start
Install Dosing Devices for automatic addition
1. Connect Dosing Devices (possible: 700/800 Dosino , 685/805
Dosimat) to the 797 VA Computrace stand
or the 846 Dosing
Interface
(see Installation of Dosing Devices, section 1.3).
2. Make hardware settings for Dosing Devices.
3. Select
Automatic
for
Addition
in the
WORKING METHOD SPECI-
FICATIONS
window.
4. Define the addition or predose solution in the
DOSINOS
win-
dow (see Dosing Devices, section 5.2).
5. The number defined in the
No.
field of the
Substances
tab of
the
EDIT WORKING METHOD PARAMETERS
window should
correspond with the number of the Dosing Device used for the
addition of this substance.
Switch on the instruments and start program
1. Switch on PC.
2. Connect Dosing Devices ( see Installation of Dosing Devices,
section 1.3), Sample Changer (see Installation of 863 Compact
VA Autosampler section 1.3, or Installation of 838 Advanced
Sample Processor section 1.3) and the 843 Pump Station to the
797 VA Computrace stand.
3. Connect the 797 VA Computrace stand to the PC (via USB).
4. Switch on 797 VA Computrace stand, Sample Changer and 731
Relay Box.
5. Start 797 VA Computrace Software 1.3.x (see Starting the VA
Computrace program, section 2.2).
6. Enter
Name
and
Password
in the
VA COMPUTRACE LOGIN
win-
dow (see Login, section 2.6).
7. Select exploratory or determination mode (see Mode menu,
section 2.4).
8. Open the desired Exploratory mode windows (Exploratory
specifications, Exploratory curves) or Determination mode win-
dows (Working method specifications, Monitor, Determination
curves, Results, Sample table).