5.3 Monitor

797 VA Computrace – Software

115

The window displays date and time of the deter-
mination start and the title of the method used.

Date and time in the format

YYMMDDHHMM

(month-day-hour-minute) and the

Sample ID

are

used as default for the name of the determination
file to be saved automatically (e.g.

0706181712_sample.dth

).

Sample ID [ 32 Zeichen ; "sample" ]

Identification for sample. The expression defined
in the

Determination

tab is displayed and can be

changed if desired.

Volume production bath (mL) [ > 0 mL ; 10 mL ]

The volume of sample bath solution added dis-
played and can be changed if desired.

Cell volume (mL) [ > 0 mL ; 10 mL ]

Total volume of solution in the measuring vessel at
the start of the determination. Is dependent on
whether you removed the electrolyte solution be-
fore you added the production bath or not.

Add the sample solution into the measuring vessel
and press

.

PLACE INTERCEPT SOLUTION

This window appears in the Electroplating Bath VA
with the

Calibration

techniques “MLAT Standard

addition for brighteners” and "LAT Record calibra-
tion curve".