5.2 Working method specifications

797 VA Computrace – Software

91

Sample identifier [ 32 characters ; "sample" ]

Identification for sample. The

Note: The

Sample identifier

is used for the file na-

me. To make sure that (for automated operations
with a sample processor) always the latest cali-
bration file is taken for the calculation; the

Sample

identifier

defined for recording the calibration cur-

ve (with

Calibration

technique "Record calibration

curve"), must match with the file name of the cali-
bration file chosen for parameter

Calibration curve

(see below) for the determination of the sample
(with

Calibration

technique "Sample with calibra-

tion curve").

Example for

Sample identifier

: CalibrationLead

Sample amount (mL) [ > 0 mL ; 10 mL ]

Amount of sample added to the measuring vessel.

Sample unit [ mL, g ; mL ]

Selection of unit for sample amount.

Note: The sample unit chosen in the

EDIT WORK-

ING METHOD PARAMETERS/Determination

tab de-

termines the selecting list for

Final Unit

in the

CAL-

CULATION

window.

mL

→ #g / #L [g/L]

g → #g / #g [g/kg]