8.6 Analyze Electroplating Bath Solutions
797 VA Computrace – Software
231
It is recommended to do one "Prep cycle".
On the
Automation
tab, select the 838 Advanced Sample Processor.
The default settings can be adopted (click button <
Default
>).
Method Parameters for the 797
It is recommended to record the "Intercept value" with the
Calibra-
tion
technique "LAT Record intercept value" (that way it is stored in
a file and can be read out automatically during the sample deter-
mination).
Therefore two methods are needed, one for the determination of
the "Intercept value", one for the determination of the sample. Set
CVS or CPVS for
Mode
for both methods.
Determination of the "Intercept value":
Choose "LAT Standard addition for brighteners" for
Calibration
technique.
In the Dosinos window, the same settings should be made as
later for the sample determination with "LAT Standard addition
for brighteners" (see below). But for the determination of the
"Intercept value" no addition is done.
Determination of the sample:
Choose "LAT Standard addition for brighteners" for
Calibration
technique.
Settings in the
DOSINOS
window: