Np – normal pulse (for "exploratory" only) – Metrohm 797 VA Computrace User Manual
Page 58
![background image](/manuals/587121/58/background.png)
3 General settings for exploratory and determination mode
797 VA Computrace – Software
46
NP – Normal Pulse (for "Exploratory" only)
General:
NP or Normal Pulse voltammetry is the classic pulse voltam-
metric measurement mode with direct recording of the current. It is
equally well suited for irreversible and reversible systems and offers
a higher sensitivity than the DC voltammetry. The NP measurement
mode can only be set for the exploratory mode by selecting
NP -
Normal pulse
for the
Mode
parameter in the
EXPLORATORY SPECI-
FICATION
window.
Description:
For NP voltammetry, square-wave pulses with an increasing ampli-
tude are superimposed on a constant base voltage. The current i is
measured as a function of the voltage U at the end of the pulse.
This normally provides wave-shaped curves which can be evaluated
using the tangent method.
U
Base potential
Voltage
step
Pulse
time
Voltage step time
t (i)
t (i)
Start potential
Half-wave voltage
Wave current
i
U