Metrohm 797 VA Computrace User Manual
Page 101
5.2 Working method specifications
797 VA Computrace – Software
89
Calibration [ see below ; Standard addition manual ]
Selection of calibration mode (see also Calibration
techniques with CVS and CPVS, section 6.2):
Standard addition
Standard addition. The number of additions is de-
fined in the
Determination
tab, the standard ad-
dition solutions are defined in the
Substances
tab,
and the
Dosing Devices
are defined in the
DOSI-
NOS
window.
Sample with calibration curve
Sample determination using previously recorded
calibration curves. The determination with the re-
corded calibration curves must be defined in the
Determination
tab.
Record calibration curve
Recording of calibration curves. The number of
additions is defined in the
Determination
tab, the
addition solutions are defined in the
Substances
tab, and the
Dosing Devices
are defined in the
DOSINOS
window.
Note: For the electroplating bath modes CVS and CPVS
the
Calibration
techniques are different. See Calibration
techniques with CVS and CPVS, section 6.2).
Technique [ see below ; Batch ]
Selection of measurement technique:
Batch
Measurement without solution exchange.
Batch with solution exchange
(not selectable with CVS and
CPVS modes)
Measurement with solution exchange for every
addition or calibration level.
Taken from calibration curve
This option is used automatically if
Sample with
calibration curve
is selected.