Measure a sample using a calibration curve, 9 work with film electrodes, Deposit a mercury film – Metrohm 797 VA Computrace User Manual
Page 264
![background image](/manuals/587121/264/background.png)
8 How to ...?
797 VA Computrace – Software
252
Measure a sample using a calibration curve
For the determination of a sample using a previously recorded ca-
libration curve, this calibration curve must have been recorded and
saved. Proceed as follows:
1. Click on
or
MAIN WINDOW
/ Mode / Determination
.
2. Click on
or
MAIN WINDOW
/ Window / Working method
specification
to open the
WORKING METHOD SPECIFICATIONS
window.
3. Load the desired method into the
WORKING METHOD SPECIFI-
CATIONS
window (see How to Load a method, section 8.4).
4. Select
Sample with calibration curve
in the
Calibration
field of
the
WORKING METHOD SPECIFICATIONS
window (Is not select-
able with the CVS and CPVS mode).
5. Click
to open the
EDIT WORKING METHOD PA-
RAMETERS
window.
6. Select
the
Determination
tab (see section 5.2) and enter
Sample
identifier
,
Sample amount
,
Cell volume
and the name and direc-
tory of the determination with the recorded calibration curve in
the
Calibration curve
field.
7. Close
the
EDIT WORKING METHOD PARAMETERS
window by
clicking
.
8. Place the sample solution in the measuring vessel at the 797
VA Computrace stand.
9. Click on
or
MAIN WINDOW
/ Window / Monitor
to open
the
MONITOR
window.
10. Start the measurement by clicking the
icon in the
MAIN
WINDOW
or the
button in the
MONITORING
win-
dow.
11. Enter the
Sample ID
(used as part of the determination file
name) in the
PLACE SAMPLE
window and click
.
8.9
Work with film electrodes
Deposit a mercury film
You find a suitable method for the determination of heavy metals
with mercury film electrodes in Application Bulletins 241 and 254.
1. Polish the glassy carbon (6.1204.110) or Ultra Trace electrode
tip (6.1204.100) with alumina powder (6.2802.000) and put
the electrode into the 797 VA Computrace stand.
2. Put
the
electrolyte
solution into the measuring vessel, e.g.:
Add 10 mL ultrapure water, 200 μL c(HCl) = 6 mol/L) and 50 μg