beautypg.com

Measure a sample using a calibration curve, 9 work with film electrodes, Deposit a mercury film – Metrohm 797 VA Computrace User Manual

Page 264

background image

8 How to ...?

797 VA Computrace – Software

252

Measure a sample using a calibration curve

For the determination of a sample using a previously recorded ca-
libration curve, this calibration curve must have been recorded and
saved. Proceed as follows:

1. Click on

or

MAIN WINDOW

/ Mode / Determination

.

2. Click on

or

MAIN WINDOW

/ Window / Working method

specification

to open the

WORKING METHOD SPECIFICATIONS

window.

3. Load the desired method into the

WORKING METHOD SPECIFI-

CATIONS

window (see How to Load a method, section 8.4).

4. Select

Sample with calibration curve

in the

Calibration

field of

the

WORKING METHOD SPECIFICATIONS

window (Is not select-

able with the CVS and CPVS mode).

5. Click

to open the

EDIT WORKING METHOD PA-

RAMETERS

window.

6. Select

the

Determination

tab (see section 5.2) and enter

Sample

identifier

,

Sample amount

,

Cell volume

and the name and direc-

tory of the determination with the recorded calibration curve in
the

Calibration curve

field.

7. Close

the

EDIT WORKING METHOD PARAMETERS

window by

clicking

.

8. Place the sample solution in the measuring vessel at the 797

VA Computrace stand.

9. Click on

or

MAIN WINDOW

/ Window / Monitor

to open

the

MONITOR

window.

10. Start the measurement by clicking the

icon in the

MAIN

WINDOW

or the

button in the

MONITORING

win-

dow.

11. Enter the

Sample ID

(used as part of the determination file

name) in the

PLACE SAMPLE

window and click

.

8.9

Work with film electrodes

Deposit a mercury film

You find a suitable method for the determination of heavy metals
with mercury film electrodes in Application Bulletins 241 and 254.
1. Polish the glassy carbon (6.1204.110) or Ultra Trace electrode

tip (6.1204.100) with alumina powder (6.2802.000) and put
the electrode into the 797 VA Computrace stand.

2. Put

the

electrolyte

solution into the measuring vessel, e.g.:

Add 10 mL ultrapure water, 200 μL c(HCl) = 6 mol/L) and 50 μg