2.7 Settings menu
797 VA Computrace – Software
29
Note: Before starting the test, switch on either 863 Compact VA
Autosampler(set Method 2) or the 838 Advanced Sample Processor
(set method VA, and enter the position of the first sample vessel
for the 838 parameter "SAMPLE" with the 838 keypad), and
place two sample vessels filled with water on the sample rack.
For details on the use of the
863 Compact VA Autosampler
, or
the
838 Advanced Sample Processor
see 863 Manual, 838
Manual.
GLP
GLP Settings
MAIN WINDOW / Settings / General Settings
The
GLP settings for the Computrace and Dosing
Devices can be defined on the
GLP
tab of the
GENERAL SETTINGS
window.
Computrace
GLP
Control
Check the checkbox to apply GLP control to the
Computrace.
Validation interval (days)
Define the period between two validations.
Display a message [ .. ] days before the validation
expires.
Define, how long before the expiring date a war-
ning message is displayed.
Action, if validation expires
You can choose between just showing a warning
Show warning only
, or stopping the measure-
ments
Stop measurements
.
Starts the GLP Wizard
Note: Clicking
resets the validation in-
terval.
Dosinos
GLP
Control
Check the checkbox to apply GLP control to the
Dosing Devices (Testing of Dosing Devices can
only be done by your Metrohm agency).
Next certification of Dosino no. 1..7 (days)
Displays days left to next certification of the spe-
cific Dosing Device.
Display a message [ .. ] days before the certifica-
tion
expires.