8.6 Analyze Electroplating Bath Solutions

797 VA Computrace – Software

237

Checking

Use for predose

for Dosino 1, VMS is added to the

measuring vessel before the sample transfer. Then Suppressor
standard solution is added via Dosino 3. That should be specified
on the

Substances

tab of the

EDIT WORKING METHOD PARAME-

TERS

window (here the window for CVS):

Determination of the sample:

Choose "DT Suppressors with calibration curve" for

Calibration

technique.
Settings in the

DOSINOS

window:

Checking

Use for predose

for Dosino 1, VMS is added to the

measuring vessel before the sample transfer. Then sample is
added via Dosino 3. That should be specified on the

Substances

tab of the

EDIT WORKING METHOD PARAMETERS

window (here

the window for CVS):