8.6 Analyze Electroplating Bath Solutions
797 VA Computrace – Software
237
Checking
Use for predose
for Dosino 1, VMS is added to the
measuring vessel before the sample transfer. Then Suppressor
standard solution is added via Dosino 3. That should be specified
on the
Substances
tab of the
EDIT WORKING METHOD PARAME-
TERS
window (here the window for CVS):
Determination of the sample:
Choose "DT Suppressors with calibration curve" for
Calibration
technique.
Settings in the
DOSINOS
window:
Checking
Use for predose
for Dosino 1, VMS is added to the
measuring vessel before the sample transfer. Then sample is
added via Dosino 3. That should be specified on the
Substances
tab of the
EDIT WORKING METHOD PARAMETERS
window (here
the window for CVS):