Metrohm 797 VA Computrace User Manual

Page 168

background image

6 Electroplating Bath VA

797 VA Computrace – Software

156

Note: If you set a small value (e.g. < 1%), it can
take a long time until reaching the target.

Note: If both

No. of conditioning measurements

[

1...100 ; 1 ] and

Auto Std.dev

(%) [ >0.1 ; 1 ] are

activated, the conditioning is finished as soon as
the first of the two criterias is reached.

No. of replications [ 1...10 ; 2 ]

Number of replications (= total number of measu-
rements) for each variation (sample, standard ad-
dition, calibration level, conditioning measure-
ment). For cyclic modes (CV, CVS, CPVS) the “total
number of measurements” is the

No. of replica-

tions

multiplied with the number of

Save last ..

sweeps

(

Voltammetric

tab). “Total number of mea-

surements” must not exceed 10.

Voltammetric tab with Calibration technique “LAT Record
intercept value”

The

Voltammetric

tab of the

EDIT WORKING METHOD PARAME-

TERS

window contains parameters for preparation procedures and

VA measurement modes. The parameters displayed depend on the
measurement mode selected in the

WORKING METHOD SPECIFICA-

TIONS

window. With “LAT Record intercept value” following pa-

rameters have to be defined:

Initial mixing time (s) [ 0...80600 s ; 10 s ]

(see

Initial mixing time

with CVS and CPVS

, section 6.3)

Time of mixing before the first measurement of
the sample solution.

Conditioning Cycles

(see

Conditioning cycles with CVS and CPVS

,

section 6.3)

Start potential (V) [ -5...+5 V ; 0.2 V ]

Start voltage for the cyclic conditioning sweep.

End potential (V) [ -5...+5 V ; 1.625 V ]

Final voltage for the cyclic conditioning sweep.

No. of cycles [ 0...X ; 0 ]

Number of conditioning cycles.

Pretreatment

(see

Pretreatment with CVS and CPVS

, section 6.3)

Cleaning potential (V) [ -5...+5 V ; 1.625 V ]

Voltage applied to the electrodes during the

Cleaning time

.