beautypg.com

2 va measurement modes, Dp – differential pulse – Metrohm 797 VA Computrace User Manual

Page 52

background image

3 General settings for exploratory and determination mode

797 VA Computrace – Software

40

3.2

VA measurement modes

DP – Differential Pulse

General:
DP or Differential Pulse voltammetry is the most universal
and frequently used voltammetric measurement mode. It is equally
well suited for irreversible and reversible systems and offers a high
sensitivity. The DP measurement mode can be set for the ex-
ploratory and determination mode by selecting

DP - Differential

pulse

for the

Mode

parameter in the

EXPLORATORY SPECIFICATION

or

WORKING METHOD SPECIFICATIONS

window.

Description:
For DP voltammetry, rectangular pulses with a constant amplitude
are superimposed on a stepwise rising direct voltage ramp. The cur-
rent i is measured as a function of the voltage U immediately be-
fore the pulse and at the end of the pulse. From the differences be-
tween the two current measurements, peak-shaped curves are ob-
tained which are evaluated using

linear, polynomial, horizontal or

exponential baselines.

U

i

U

Start potential

Pulse

amplitude

Voltage step

Pulse

time

Voltage step time

Peak voltage

Peak

height

t

i = i2 - i1

t (i1) t (i2)

t (i1) t (i2)