Substance evaluation, Peak evaluation, Substance evaluation peak evaluation – Metrohm 797 VA Computrace User Manual
Page 143

5.5 Results
797 VA Computrace – Software
131
Note: With the electroplating bath modes CVS and CPVS, and the
Calibration
technique "LAT Record intercept value"
Volume Inter-
cept Solution
is indicated instead of
Sample amount
.
Note: With the electroplating bath modes CVS and CPVS, and the
Calibration
techniques "LAT Standard addition for brighteners",
"MLAT Standard addition for brighteners" and "RC Sample with re-
sponse curve"
Volume prod. bath
is displayed instead of
Sample
amount
.
Substance evaluation
Substance
Substance name (see
Substance
on the
Substances
tab,
section 5.2).
Conc.
Calculated substance concentration referring to the total
volume of solution (
Cell volume
) in the measuring vessel at
the start of the first sweep.
Conc.dev.
Absolute and relative total deviation of the calculated sub-
stance concentration
Conc.
.
Amount
Absolute substance amount in the measuring vessel.
Add.amount
Substance amount added in every standard addition (only
available for constant additions).
Note: For CVS and CPVS, additional evaluation values are dis-
played.
Peak evaluation
VR
Number of variation and replication.
V
Evaluated peak voltage (V).
A;W;A/V;C {for PSA/CCPSA: s/V;s;s/(V*V)}
Calculated eval. quantity (height, area, derivative or
charge)
i.mean; P.mean; mean; Q.mean(only with modes CVS and
CPVS)
Mean value of the evaluation quantity for all replications of
a variation.
Std.Dev.
Standard deviation of the evaluation quantity for all repli-
cations of a variation.