8 How to ...?

797 VA Computrace – Software

250

1. Click on

or

MAIN WINDOW

/ Mode / Determination

.

2. Click on

or

MAIN WINDOW

/ Window / Working method

specification

to open the

WORKING METHOD SPECIFICATIONS

window.

3. Load the desired method into the

WORKING METHOD SPECIFI-

CATIONS

window (see How to Load a method, section 8.4).

4. Select

Record calibration curve

in the

Calibration

field of the

WORKING METHOD SPECIFICATIONS

window (Is not selectable

with the CVS and CPVS mode).

5. Select

Manual

in the

Addition

field of the

WORKING METHOD

SPECIFICATIONS

window.

6. Select

Batch with solution exchange

in the

Technique

field of

the

WORKING METHOD SPECIFICATIONS

window.

7. Click

to open the

EDIT WORKING METHOD PA-

RAMETERS

window.

8. Select

the

Determination

tab (see section 5.2) and enter the

Cell

volume

and the number of calibration solutions in the

No. of

cells

field.

9. Select

the

Substances

tab (see section 5.2) and make sure that

for each substance entered in the table the concentrations of
the calibration solutions are defined in the

CELL CONCEN-

TRATIONS

window which is opened by clicking on the

button.

10. Close the

EDIT WORKING METHOD PARAMETERS

window by

clicking

.

11. Place the first calibration solution in the measuring vessel at the

797 VA Computrace stand.

12. Click on

or

MAIN WINDOW

/ Window / Monitor

to open

the

MONITOR

window.

13. Start the measurement by clicking the

icon in the

MAIN

WINDOW

or the

button in the

MONITORING

win-

dow.

14. Enter the

Calibration curve id

(used as part of the determination

file name) in the

START CALIBRATION

window and click

.

15. Each time a solution exchange is required in the

BATCH SOLU-

TION EXCHANGE

window, replace the solution measured by

the next calibration solution.

Record calibration curve automatically

This method is used for preparing different calibration solutions by
adding automatically several times a concentrated standard solu-