11. Start the measurement by clicking the

icon or the

button (see Performing exploratory measurements,

section 4.2).

Note: As soon as a signal curve is measured or loaded, the volt-
ammetric parameter get read only. Should the parameters be
changed before measuring a new signal curve, the button

needs to be clicked.

Evaluate signal peaks automatically

1. Select the desired signal file in the

Signal

field of the

EXPLORA-

TORY CURVES

window. The selected signal curve is shown with

the

Selected signal properties

.

2. Click

on

EXPLORATORY CURVES

/ Signal / Peak search

. The

PEAK SEARCH

window is opened.

3. Set the parameters

Reverse peak

,

Reverse sweep

(for CV and

CVS only),

Minimum peak width

,

Smooth factor

,

Minimum peak

height

and

Scope

for peak evaluation (see Peak search, section

4.3).

4. Click the

button. The calculated baselines and

peak maximum positions are displayed in the

EXPLORATORY

CURVES

window. The evaluation results are displayed in the ta-

ble of results in the

PEAK SEARCH

window.

5. If no peaks are found, try to modify the peak search parame-

ters

Minimum peak width

,

Smooth factor

, and

Minimum peak

height

for the automatic signal peak evaluation or switch to the

manual signal peak evaluation.

Evaluate signal peaks manually

1. Select the desired signal file in the

Signal

field of the

EXPLORA-

TORY SPECIFICATION

window. The selected signal curve is

shown with the

Selected signal properties

.

2. Click

on

EXPLORATORY CURVES

/ Signal / Peak search

. The

PEAK SEARCH

window is opened.

3. Check

the

Manual

option.

4. Set the parameters

Reverse peak

,

Reverse sweep

(for CV and

CVS only),

Minimum peak width

,

Smooth factor

, and

Minimum

peak height

for peak evaluation (see Peak search, section 4.3).

5. Set the start and end base points for baseline evaluation by

clicking the

buttons of the

Begin

or

End

field.

6. Select

Type

and

Scope

of the baseline.