Load/save sample table – Metrohm 797 VA Computrace User Manual
Page 147
5.6 Sample table
797 VA Computrace – Software
135
Sample ID [ 32 characters ; "sample" ]
Identification for sample.
Amount [ > 0 ; 10 ]
Amount of sample added to the measuring vessel.
Unit [ mL, g ; mL ]
Selection of unit for sample amount.
Cell volume (mL) [ > 0 mL ; 10 mL ]
Total volume of solution (sample + auxiliary solu-
tion, e.g. buffer) in the measuring vessel at the
start of the determination. The sample concentra-
tions
Conc.
refer to this cell volume.
Method
Selection of the method used for this sample.
Status [ read only ]
Display of sample determination status:
Measur-
ing
,
Done
or "empty" (ready for start).
Add a new sample data row. The
SAMPLE
win-
dow for entry of
Method
,
Sample ID
,
Sample
amount
,
Sample unit
, and
Cell volume
is opened.
Edit the selected sample data row. The
SAMPLE
window for modification of
Method
,
Sample ID
,
Sample amount
,
Sample unit
, and
Cell volume
is
opened.
Reset
the
Status
of the sample data rows to "emp-
ty" in order to restart the current sample table.
Clear the contents of the selected sample data
row.
Print the sample table.
Load/save sample table
Sample table files (
*.spt
) containing sample table data can be cre-
ated, loaded and saved by the following commands:
SAMPLE TABLE
/ File / New
Load an empty sample table into the working
memory.
SAMPLE TABLE / File / Load
Load an existing sample table into the working
memory. The name of the sample table is dis-
played in the status bar of the
MAIN WINDOW
.
SAMPLE TABLE
/ File / Save
Save the current sample table loaded in the wor-
king memory. If the sample table has been chan-
ged since loading, the message
The file already
exists. Overwrite?
appears. Click
Yes
to overwrite