3.2 VA measurement modes

797 VA Computrace – Software

43

Sweep parameters:

Hydrodynamic (measurement) [ on, off ; off ]

Enable/disable stirring of the RDE/SSE during the sweep.

Start potential (V)

[ -5...+5 V ; -0.9 V ]

Start voltage for the voltage sweep.

End potential (V)

[ -5...+5 V ; -0.1 V ]

Final voltage for the voltage sweep.

Voltage step (V)

[ > 0 V ; 0.006 V ]

Voltage step for direct voltage ramp.

Amplitude (V)

[ > 0...+1 V ; 0.02 V ]

Voltage amplitude of the square wave voltage superimposed
on the direct voltage.

Frequency (Hz)

[ > 0...2000 Hz ; 50 Hz ]

Frequency of the superimposed square wave voltage, which
defines the voltage step time (

Voltage step time = 1 / Fre-

quency

).

Sweep rate (V/s)

[ read only ]

Display of the ramp slope calculated as

Voltage step × Fre-

quency

.

Notes:
• The SqW measurement mode can only be used with HMDE or

RDE electrodes.

If your curves have high noise, put

Highest current range

and

Lowest current range

in the

POTENTIOSTAT

window on the

same level (see section 9.6

SqW Problems

)

• The following condition applies to the definition of the

Voltage

step time

:

Frequency 2000 Hz Voltage step time 500 µs

• The measurement time

t (i)

is defined as follows:

Voltage step time 80 ms

t (i) = 20/16.67 ms

(power fre-

quency 50/60 Hz)

Voltage step time < 80 ms

t (i) = 0.5 × Voltage step time