5 Determination mode

797 VA Computrace – Software

108

Use for predose [ on, off ; off ]

Use of a

Dosing Device

for addition of an auxiliary

solution (e.g. buffer, or VMS for automated elec-
troplating bath analysis) at the start of the de-
termination.

Volume of predose (mL) [ > 0.01 mL ; 0 mL ]

Volume of the auxiliary solution (e.g. buffer, or
VMS for automated electroplating bath analysis)
to be added at the start of the determination.

This

volume must be accounted for the calculation of
the

Cell volume

.

Use after sample transfer [ on, off ; off ]

(only with the to electro-

plating bath modes CVS and CPVS and the

Cali-

bration

technique "MLAT" or "RC Sample with re-

sponse curve")

Use of a Dosing Device for addition of an auxiliary
solution (e.g. Suppressor) after the sample trans-
fer.

Volume after sample transfer [ 0.01 ... 1000 mL ; 0 mL ]

(only with

the two electroplating bath modes CVS and CPVS
and the

Calibration

technique "MLAT")

Volume for addition of an auxiliary solution (e.g.
Suppressor) after the sample transfer.

Content [ 46 characters ; "solution" ]

Remarks regarding the solution.

Note: Dosing Device parameters can only be overwritten if the
checkbox

Use

is checked.

5.3 Monitor

Start determination

If no determination is running, the

MONITOR

window is used to

start a new determination with the current working method.