3 monitor, Start determination – Metrohm 797 VA Computrace User Manual
Page 120
5 Determination mode
797 VA Computrace – Software
108
Use for predose [ on, off ; off ]
Use of a
Dosing Device
for addition of an auxiliary
solution (e.g. buffer, or VMS for automated elec-
troplating bath analysis) at the start of the de-
termination.
Volume of predose (mL) [ > 0.01 mL ; 0 mL ]
Volume of the auxiliary solution (e.g. buffer, or
VMS for automated electroplating bath analysis)
to be added at the start of the determination.
This
volume must be accounted for the calculation of
the
Cell volume
.
Use after sample transfer [ on, off ; off ]
(only with the to electro-
plating bath modes CVS and CPVS and the
Cali-
bration
technique "MLAT" or "RC Sample with re-
sponse curve")
Use of a Dosing Device for addition of an auxiliary
solution (e.g. Suppressor) after the sample trans-
fer.
Volume after sample transfer [ 0.01 ... 1000 mL ; 0 mL ]
(only with
the two electroplating bath modes CVS and CPVS
and the
Calibration
technique "MLAT")
Volume for addition of an auxiliary solution (e.g.
Suppressor) after the sample transfer.
Content [ 46 characters ; "solution" ]
Remarks regarding the solution.
Note: Dosing Device parameters can only be overwritten if the
checkbox
Use
is checked.
5.3 Monitor
Start determination
If no determination is running, the
MONITOR
window is used to
start a new determination with the current working method.