7 Manual control
797 VA Computrace – Software
204
Purge time (s)
[ 0...80600 s ; 300 s ]
Time of inert gas purging before the first measurement of the
sample solution.
Conditioning
cycles
Electrochemical regeneration of the solid state electrode by a
freely selectable number of conditioning cycles. For every cy-
cle, the voltage is changed at a sweep rate of 1 V/s to the
End
potential
and then decreased at the same rate back to the
Start potential
.
Start potential (V)
[ -5...+5 V ; -1.2 V ]
Start voltage for the cyclic conditioning sweep.
End potential (V)
[ -5...+5 V ; -0.1 V ]
Final voltage for the cyclic conditioning sweep.
No. of cycles
[ 0...X ; 0 ]
Number of conditioning cycles.
Cleaning potential (V)
[ -5...+5 V ; -0.8 V ]
Voltage applied to the electrodes during the
Cleaning time
.
Cleaning time (s)
[ 0...80600 s ; 10 s ]
Time during which the
Cleaning
potential
is applied to the
electrodes.
Sweep
Parameters of DP sweep used at the end of the cleaning cyc-
les for checking the electrode (see VA measurement modes,
section 3.2).
No. of repetition cycles
[ 0...X ; 5 ]
Number of repetition cycles for applying the conditioning cyc-
les and cleaning potential steps.
Cell off after measurement
[ on, off ; on ]
Enable/disable the switching off of the voltage applied to the
electrodes after measurement.
Stand-by potential (V)
[ -5...+5 V ; -0.1 V ]
Voltage to be applied to the electrodes after measurement if
the
Cell off after measurement
box is set to
off
.