5 Determination mode

797 VA Computrace – Software

94

For a detailed description of these parameters see section 3.4 Gen-
eral operation sequence
, and section 3.2 VA measurement modes.

If you work with the electroplating bath modes

CVS

or

CPVS

, see

section 8.6

Operate a sequence in Electroplating Bath VA

.

Note: For

Calibration

technique "Sample with calibration curve"

the

Voltammetric

tab parameters are not editable. To make sure

that the

Voltammetric

parameters defined for the determination of

the sample match with the

Voltammetric

parameters defined for re-

cording the calibration curve (with

Calibration

technique "Record

Calibration curve"), following procedure is recommended:

1. Load the determination of the calibration curve (

FILE / Load de-

termination

)

2. Open window

DETERMINATION CURVES

3. Copy the parameters to the working method (

EDIT / Copy pa-

rameters to working method

)

The same procedure is recommended for the Suppressor determi-
nation (see

Voltammetric

tab with

Calibration

technique "DT Supp-

ressors with calibration curve").

Substances

The

Substances

tab of the

EDIT WORKING METHOD PARAMETERS

window contains parameters for the definition and recognition of
substances, for the definition of standard solutions, for peak
evaluation and results calculation. T

he parameters displayed de-