8 How to ...?

797 VA Computrace – Software

242

Es wird für beide Dosino ein "Prep-Zyklus" empfohlen.

On the

Automation

tab, select the 838 Advanced Sample Processor.

The default settings can be adopted (click button <

Default

>).

Method Parameters for the 797
Two steps are required for Suppressor determination. One to re-
cord the response curve with

Calibration

technique "RC Record re-

sponse curve", another for the actual sample determination with

Calibration

technique "RC Sample with response curve". Set CVS or

CPVS for

Mode

for both methods.

Recording of the

Response Curve:

:

Choose "RC Record response curve" for

Calibration

.

Settings in the Dosinos window:

After the addition of Electrolyte solution, Suppressor standard
solution is added via Dosino 3. That should be specified on the

Substances

tab of the

EDIT WORKING METHOD PARAMETERS

window (here the window for CVS):