8 How to ...?

797 VA Computrace – Software

248

8. Check the Dosing Devices, which are used for standard addi-

tion in the

Use

field (see Dosing Devices, section 5.2)

9. Close

the

DOSINOS

window by clicking

.

10. Click

to open the

EDIT WORKING METHOD PA-

RAMETERS

window.

11. Select the

Determination

tab (see section 5.2) and enter

Sample

identifier

,

Sample amount

,

Cell volume

, and the number of

standard additions in the

No. of additions

field (If you work

with the CVS or CPVS mode, see section 6.2 Calibration tech-
niques with CVS and CPVS
for a description of the parameters
on the

Determination

tab).

12. Select the

Substances

tab (see section 5.2) and make sure that

for each substance entered in the table the number of the sin-
gle or mixed standard addition solution, its concentration and
its volume is defined. The

No.

of the standard solution must be

identical to the number of the Dosing Devices used for auto-
matic addition of this solution.

13. If the standard addition should be done with variable addition

volumes, click the button in the

Volume

column to open

the

EDIT VARIED ADDITION

window, enter the variable addi-

tion volumes in the

Addition

fields, and close this window by

clicking

.

14. Close the

EDIT WORKING METHOD PARAMETERS

window by

clicking

.

15. Place the sample solution in the measuring vessel at the 797

VA Computrace stand.

16. Click on

or

MAIN WINDOW

/ Window / Monitor

to open

the

MONITOR

window.

17. Start the measurement by clicking the

icon in the

MAIN

WINDOW

or the

button in the

MONITORING

win-

dow.

18. Enter the

Sample ID

(used as part of the determination file

name) in the

PLACE SAMPLE

window and click

.

8.8

Calibration curve technique

Record calibration curve manually by adding standard solution

This method is used for preparing different calibration solutions by
adding several times a concentrated standard solution to the meas-
uring solution using a pipette. Proceed as follows:

1. Click on

or

MAIN WINDOW

/ Mode / Determination

.