8.10 Diagnostic procedures

797 VA Computrace – Software

253

c(Hg(II)) = 1 g/L to the empty measuring vessel at the 797 VA
Computrace stand.

3. Click

on

MAIN WINDOW

/ Utility / Film deposition

to open the

FILM DEPOSITION

window.

4. Enter suitable parameters in the parameter list.
5. Click on the

button.

6. Check the resulting test voltammogram. The voltammogram

should show low noise and a low background current (low μA
range). No interfering peaks should be visible.

Remove a mercury film

Mercury films can be easily wiped off manually with a tissue. The
cleaning procedure can be used instead to remove the mercury film
electrochemically or to clean the electrode surface after having re-
moved the mercury film mechanically.
1. Put the cleaning solution into the measuring vessel, e.g.:

Add 10 mL ultrapure water and 1 mL w(HNO

3

) = 0.65 to the

empty measuring vessel at the 797 VA Computrace stand.

2. Click

on

MAIN WINDOW

/ Utility / Cleaning procedure

to open

the

CLEANING PROCEDURE

window.

3. Enter suitable parameters in the parameter list.
4. Click on the

button.

5. Check the resulting test voltammogram. The resulting voltam-

mogram should show low noise and a low background current
(low μA range). If all mercury had been oxidized, no mercury
peak remains.

8.10 Diagnostic

procedures

Check the purging

1. Connect the inert gas to the 797 VA Computrace stand (see

Hardware Manual).

2. Make sure that the inert gas pressure is 1 ± 0.2 bar.
3. Add 20 mL ultrapure water to the empty measuring vessel at

the 797 VA Computrace stand.

4. Click on

or

MAIN WINDOW

/ Utility / Computrace control

to open the

COMPUTRACE CONTROL

window.

5. Select

HMDE

and click on

.

6. Make sure that inert gas bubbles are purging through the solu-

tion.