Metrohm 797 VA Computrace User Manual
Page 236
8 How to ...?
797 VA Computrace – Software
224
The “Addition mixing time” can be defined on the
Deter-
mination
tab of the
EDIT WORKING METHOD PARAME-
TERS
window.
6. Second Measurement:
a.
Conditioning cycles
(Number of “conditioning cycles” de-
fined in
No. of cycles
on the
Voltammetric
tab of the
EDIT
WORKING METHOD PARAMETERS
window)
b.
Pretreatment
c.
Sweep
(Number of sweeps is defined in
No. of sweeps
,
number of saved sweeps in
Save last .. sweeps
on the
Volt-
ammetric
tab of the
EDIT WORKING METHOD PARAME-
TERS
window)
Note: Instead of repeating sweeps, you can also repeat
the whole Measurement cycle. Define the number of repli-
cations in
No. of replications
on the
Determination
tab of
the
EDIT WORKING METHOD PARAMETERS
window. Met-
rohm recommends to set the
No. of replications
on 1, the
No. of sweeps
on 4 and
Save last .. sweeps
on 2.
7. Second Addition
8.
Addition mixing time
Continue with measurement, addition, mixing as long as necessary.
Brightener Analysis with 838 Advanced Sample Processor and "MLAT"
Following installations and settings are recommended for the au-
tomated Brightener determination with the 838 Advanced Sample
Processor and
Calibration
technique "MLAT":
Instruments
Install the
838 Advanced Sample Processor
, three
800 Dosi-
nos
and a
732 Relay Box
with two
823 Membrane Pump
Units
(see Hardware Manual 797 and Instructions for Use 838).
Dosinos:
Dosino 1:
50 mL Exchange Unit
VMS solution
Dosino 2:
2 mL Exchange Unit
Brightener standard so-
lution
Dosino 3:
2 mL Exchange Unit
Suppressor concentrate
Method at 838
Before each start, define the position of the first sample vessel for
the 838 parameter "
SAMPLE
" with the 838 keypad.
Set method
LAT
at the 838 Advanced Sample Processor.