beautypg.com

Cvs - cyclic voltammetric stripping – Metrohm 797 VA Computrace User Manual

Page 67

background image

3.2 VA measurement modes

797 VA Computrace – Software

55

Voltage step time > Modulation time + 450 ms

(HMDE/RDE)

Voltage step time > Modulation time + 470 ms

(DME)

Voltage step time > Modulation time + Drop size * 40 ms + 450
ms

(SMDE)

Modulation time > 2 / Frequency

• The measurement time

t (i)

is defined as follows:

t (i) = Modulation time / 2

CVS - Cyclic Voltammetric Stripping

General:
CVS

or

Cyclic Voltammetric Stripping

is used to determine

organic additives in electroplating electrolytes. The CVS measuring
mode can be set for the exploratory and determination mode by
selecting

CVS - Cyclic Voltammetric Stripping

for the

Mode

parame-

ter in the

EXPLORATORY SPECIFICATION

or

WORKING METHOD

SPECIFICATIONS

window

.

Description:

For Cyclic Voltammetric Stripping, the potential of an RDE (e.g. Pt)
is cycled at a constant rate in a plating bath. The current

i

is meas-

ured as a function of the voltage

U

. The curves registered in the

last cycles are stored and its peaks can be evaluated using linear,
polynomial, horizontal or exponential baselines.

i

an

U

i

cath

U

t

Conditioning

Plating

Stripping

Cleaning

Cleaning

Conditioning

Stripping

Plating