8.6 Analyze Electroplating Bath Solutions

797 VA Computrace – Software

229

Instruments
Install the 838 Advanced Sample Processor, a 800 Dosino, a
732 Relay Box with two 823 Membrane Pump Units (see
Hardware Manual 797 and 838 Instructions for Use).

Dosinos:

(Dosino 1:

not used)

Dosino 2:

2 mL Exchange Unit

Brightener standard so-
lution

(Dosino 3:

not used)

Method at 838

Before each start, define the position of the first sample vessel for
the 838 parameter "

SAMPLE

" with the 838 keypad.

Set method

LAT

at the 838 Advanced Sample Processor.

LAT

parameters

method LAT
number of samples

rack

> Number of samples to be

processed (entire sample
rack)

>start sequence

1 CTL:Rm:

INIT

> Initialize Remote lines

2 Move

1

:

sample

> First Sample on Pos. "SAM-

PLE"

3 CTL:Rm:

************1

> Remote start CT797

4 CTL:Rm:

************0

>sample sequence

1 SCN:Rm:

*****1**

> Scan line for incoming signal

from CT797

2 MOVE

1

:

sample

> Move needle to sample

3 LIFT: 1

:

work mm

> Place needle at working posi-

tion

4 PERISTALT: 300 s

10

> Sample

transfer

5 CTL:Rm:

************1

> Set line: sample transfer

ready

6 CTL:Rm:

************0

7 MOVE

1

:

+28

> Move needle to rinsing posi-

tion

8 LIFT: 1

work mm

> Place needle at working posi-

tion

9 PERISTALT: 5 s

10

> Suck rinsing solution in the

needle

10 SCN:Rm:

*****1**

> Scan line for incoming signal

from CT797

11 PERISTALT:

300 s

10

> Rinse

needle

12 CTL:Rm:

************1

> Set line: needle cleaning

ready

13 CTL:Rm:

************0

Samples