Metrohm 797 VA Computrace User Manual

Page 196

background image

6 Electroplating Bath VA

797 VA Computrace – Software

184

Sample identifier [ 32 characters ; "sample" ]

Identification for sample.

Note: The

Sample identifier

is used for the file na-

me. To make sure that for "Suppressor analysis
with 838 Advanced Sample Processor and RC" al-
ways the latest response curve file is taken for the
calculation, this

Sample identifier

must match with

the name of the calibration file defined for the pa-
rameter

Response curve

on the

Determination

tab

with calibration "

RC Record response curve

".

Example for

Sample identifier

:

Response Suppressor

Note: To make sure that for "Suppressor analysis
with 838 Advanced Sample Processor and RC" al-
ways the latest response curve file is taken for the
calculation, the path of the calibration file defined
for the parameter

Response curve

on the

Sub-

stances

tab with

Calibration

technique "RC Sample

with response curve”, must match with the path
defined for the parameter

Data folder

(for the cur-

rently logged-in user) on the tab

User Directories

of the window

USER RIGHTS

.

Note: A determined response curve file is only
adopted automatically if the checkbox

Save cali-

bration curves additionally without date and time

on the

General

tab of the

GENERAL SETTINGS

window is checked.

Volume electrolyte (mL) [ > 0 mL ; 10 mL]

The volume of the Electrolyte Solution (see

Electrolyte solution

, section 6.4

).

Cell volume (mL) [ > 0 mL ; 10 mL ] [ read only ]

Total volume of solution (equal to the volume of
the Electrolyte solution) in the measuring vessel at
the start of the determination. The sample con-
centrations

Conc

. calculated refers to this cell vol-

ume.

Initial electrode conditioning

(

see Initial electrode conditioning,

section 6.3

)

Check the checkbox to use conditioning.

No. of conditioning measurements [ 1...100 ; 1 ]

Number of replications of conditioning measure-
ments. The “total number of conditioning mea-
surements” is the

No. of conditioning measu-

rements

multiplied with the number of

Save last ..

sweeps

(

Voltammetric

tab). “Total number of con-

ditioning measurements” must not exceed 100.