Metrohm 797 VA Computrace User Manual
Page 63
3.2 VA measurement modes
797 VA Computrace – Software
51
Sweep parameters:
Hydrodynamic (measurement) [ on, off ; off ]
Enable/disable stirring of the RDE/SSE during the sweep.
Potential limit (V)
[ -5...+5 V ; -0.2 V ]
Voltage limit for PSA sweep.
Maximum time (s)
[ > 0 ; 20 s ]
Time limit for PSA sweep.
Notes:
• The PSA measurement mode should only be used with RDE
electrodes (mainly with Hg film).
CCPSA – Constant Current Potentiometric Stripping Analysis
General:
CCPSA
or
Constant Current Potentiometric stripping
analysis
with oxidation by an applied constant current is mainly
used to determine substances in an organic matrix with the aid of
mercury film electrodes without prior digestion. The CCPSA measu-
rement mode can be set for the exploratory and determination mo-
de by selecting
CCPSA – Constant Current Potentiometric stripping
analysis
for the
Mode
parameter in the
EXPLORATORY SPECIFICA-
TION
or
WORKING METHOD SPECIFICATIONS
window.
Description:
In CCPSA measurement mode, the analytes are deposited at the
working electrode with the constant
Deposition potential
during a
predetermined
Deposition time
. Then the applied deposition poten-
tial is switched off and a constant current is applied. The voltage U
is measured as a function of the time t with a sampling rate of
69.69 kHz. The measurement time is limited either by the set
Po-
tential limit
or the
Maximum time
. The voltage measurement
U vs t
is used to calculate the retention times dt/dU vs U. This results in
peak-shaped curves which can be evaluated. The
Peak voltage
is
characteristic of the substance, the
Peak area
is proportional to its
concentration.