8 How to ...?

797 VA Computrace – Software

230

50 mL sample vessels can be placed on the two outer rings (11 mL
sample vessels on the two inner rings).

Arrangement: Place on the outer of the two rings alternately "In-
tercept solution" and sample solution, on the inner of the two rings
the rinsing solution.

Example for an arrangement with 7 samples:

Intercept solution

Rinsing solution

not used

Sample

1

2

3

28

27

29

30

31

56

55

57

58

59

84

83

85

86

87

112

111

Note: Although the first sample is placed on position 2, position
1 remains the starting position in the start sequence of the
method at the 838 (recording the "Intercept value").

Note:

Before

each

start, define the position of the first sample

vessel for the 838 parameter "

SAMPLE

" with the 838 keypad.

General Settings in the 797 Software
Activate both options on the

General

tab of the

GENERAL SET-

TINGS

window, so that a recorded "Intercept value" replaces the

old one automatically.
Settings on the

Dosinos

tab of the

GENERAL SETTINGS

window in

the 797 Software: