8 How to ...?
797 VA Computrace – Software
230
50 mL sample vessels can be placed on the two outer rings (11 mL
sample vessels on the two inner rings).
Arrangement: Place on the outer of the two rings alternately "In-
tercept solution" and sample solution, on the inner of the two rings
the rinsing solution.
Example for an arrangement with 7 samples:
Intercept solution
Rinsing solution
not used
Sample
1
2
3
28
27
29
30
31
56
55
57
58
59
84
83
85
86
87
112
111
Note: Although the first sample is placed on position 2, position
1 remains the starting position in the start sequence of the
method at the 838 (recording the "Intercept value").
Note:
Before
each
start, define the position of the first sample
vessel for the 838 parameter "
SAMPLE
" with the 838 keypad.
General Settings in the 797 Software
Activate both options on the
General
tab of the
GENERAL SET-
TINGS
window, so that a recorded "Intercept value" replaces the
old one automatically.
Settings on the
Dosinos
tab of the
GENERAL SETTINGS
window in
the 797 Software: