6 Electroplating Bath VA

797 VA Computrace – Software

178

this case,

var

is entered into the field instead of a

fixed value. This field only appears once for soluti-
ons with the same number (mixed standards) and
it is not displayed if

0

is entered as solution num-

ber.

Regression technique [ see below ; Linear Regression ]

Selection of regression technique:

Linear

Regression

The regression is calculated with a straight line.

Nonlinear

Regression

The regression is calculated with a nonlinear
curve.

Linear

Interpolation

The regression is calculated with a linear interpola-
tion through 2 points.

Quadratic

Regression

The regression is calculated with a nonlinear
curve.

Peak evaluation [ Height, Area, Derivative, Coulometric ; Coulomet-

ric ]

Selection of peak evaluation quantity:

Height

Peak height from baseline to peak maximum.

Area

Peak area between peak curve and calculated ba-
seline.

Derivative

Difference between positive an negative maximum
of the first derivative of the voltammogram.

Coulometric

Charge transmitted during the peak.

Smooth factor [ 1...6 ; 4 ]

Smoothing power of the Savitzky/Golay smooth-
ing of the baseline (1 = minimum smoothing, 6 =
maximum smoothing) (s

ee

Smoothing and differ-

entiation

, section 5.8

).

Eliminate spikes

Eliminates spikes to smooth the signal.

Reverse sweep [ on, off ; off ]

Enable peak evaluation of the reverse sweep of
cyclic sweeps (not available with CPVS).

Endpoint and evaluation criteria:

Addition ratio Q/Q(0) [ 0…1 ; 0.3]

The „Addition ratio“ defines the endpoint criteri-
on. After reaching this point, the standard additi-