Metrohm 797 VA Computrace User Manual
Page 107

5.2 Working method specifications
797 VA Computrace – Software
95
pend on the
Calibration
technique selected in the
WORKING
METHOD SPECIFICATIONS
window.
Substance [ 24 characters ; ]
Substance name. For the assignment of a found
peak to this substance the checkbox on the left
side of the substance name must be checked.
Peak pos. +/- (V) [ -5...+5 V ; 0 V ]
Position of the peak voltage for the substance and
tolerance for this verification voltage.
Bsln.
Parameters for baseline evaluation (details see
Baseline, section 5.2). Click the button to
open the
BASELINE
window for the selected sub-
stance.
Standard solution
(
Additive solution
with CVS and CPVS)
Definition of addition solutions for standard addi-
tion or recording of calibration curves. These pa-
rameters are not displayed if
Batch with solution
exchange
is selected for
Technique
.
No. [ 0...8; 0 ]
Number of addition solution used for manual or
automatic addition. For automatic additions, this
number is also the Dosing Device number. If a
mixed standard solution is used, the number of
this solution must be entered for each substance
present in the mixed solution.