5.2 Working method specifications

797 VA Computrace – Software

95

pend on the

Calibration

technique selected in the

WORKING

METHOD SPECIFICATIONS

window.

Substance [ 24 characters ; ]

Substance name. For the assignment of a found
peak to this substance the checkbox on the left
side of the substance name must be checked.

Peak pos. +/- (V) [ -5...+5 V ; 0 V ]

Position of the peak voltage for the substance and
tolerance for this verification voltage.

Bsln.

Parameters for baseline evaluation (details see
Baseline, section 5.2). Click the button to
open the

BASELINE

window for the selected sub-

stance.

Standard solution

(

Additive solution

with CVS and CPVS)

Definition of addition solutions for standard addi-
tion or recording of calibration curves. These pa-
rameters are not displayed if

Batch with solution

exchange

is selected for

Technique

.

No. [ 0...8; 0 ]

Number of addition solution used for manual or
automatic addition. For automatic additions, this
number is also the Dosing Device number. If a
mixed standard solution is used, the number of
this solution must be entered for each substance
present in the mixed solution.